Table 1 Optimum calibration of the SBP4 MI-ELISA using a reference serum panel

From: A novel modified-indirect ELISA based on spherical body protein 4 for detecting antibody during acute and long-term infections with diverse Babesia bovis strains

Sample, dilutiona

Mean OD

S/N ratio

Resultb

Positive serum C167, 1/64

0.679

7.5

+

Positive serum C167, 1/128

0.394

4.3

+

Positive serum C167, 1/256

0.278

3.1

+

Positive serum C167, 1/512

0.187

2.1

–

Negative serum #37

0.123

1.3

–

Negative serum #312

0.149

1.6

–

  1. Abbreviations: OD optical density, S OD of test sample, N OD of negative control
  2. aPositive serum was diluted with negative serum