Skip to main content

Table 4 Comparison of PrioCHECK ® Toxoplasma Ab SR ELISA-results (cut-off PP 20) from serum samples of 300 sheep from Austria (Groups 3 and 4) to results obtained by IFAT (cut-offs 1:40 and 1:80) and IHA and inter-rater agreement (kappa value)

From: Evaluation of a commercial ELISA kit for detection of antibodies against Toxoplasma gondii in serum, plasma and meat juice from experimentally and naturally infected sheep

  

IFAT serum cut-off 1:40

IFAT serum cut-off 1:80

IHA serum cut-off 1:80

  

positive

negative

total

positive

negative

total

positive

negative

total

ELISA serum

positive

139

2

141

135

6

141

138

3

141

 

negative

11

148

159

5

154

159

3

156

159

 

total

150

150

300

140

160

300

141

159

300

 

Kappa

0.913

  

0.926

  

0.960

  
 

Standard Error

0.0235

  

0.0218

  

0.0162

  
 

95% CI

0.867-0.959

  

0.884-0.969

  

0.928-0.992