Skip to main content

Table 3 Product test result (absolute number)

From: A comparative evaluation of the performance of commercially available rapid immunochromatographic tests for the diagnosis of visceral leishmaniasis in Bangladesh

Product Case (30)
Fresh Stored
Blood Serum Blood Serum
Positive Negative Invalid Positive Negative Invalid Positive Negative Invalid Positive Negative Invalid
Kalazar Detect 30 00 00 30 00 00 30 00 00 30 00 00
Signal®KA 30 00 00 29 01 00 30 00 00 30 00 00
Crystal®KA 30 00 00 30 00 00 30 00 00 30 00 00
Onsite leishmania Ab (Rev A) 30 00 00 30 00 00 - - - 30 00 00
Onsite leishmania Ab (Rev B) 30 00 00 30 00 00 - - - 30 00 00
Product Healthy Control (35)
Fresh Stored
Blood Serum Blood Serum
Positive Negative Invalid Positive Negative Invalid Positive Negative Invalid Positive Negative Invalid
Kalazar Detect 00 35 00 00 35 00 00 35 00 01 34 00
Signal®KA 00 34 01 00 35 00 01 34 00 00 35 00
Crystal®KA 00 35 00 00 35 00 00 35 00 00 35 00
Onsite leishmania Ab (Rev A) 01 34 00 02 33 00 - - - 00 35 00
Onsite leishmania Ab (Rev B) 18 17 00 19 16 00 - - - 00 35 00
Product Disease Control (30)
Fresh Stored
Blood Serum Blood Serum
Positive Negative Invalid Positive Negative Invalid Positive Negative Invalid Positive Negative Invalid
Kalazar Detect 02 28 00 02 28 00 01 29 00 01 28 01
Signal®KA 02 28 00 02 28 00 01 25 04 01 28 01
Crystal®KA 01 29 00 01 29 00 01 29 00 01 28 01
Onsite leishmania Ab (Rev A) 02 28 00 02 28 00 - - - 03 26 01
Onsite leishmania Ab (Rev B) 09 21 00 10 20 00 - - - 16 13 01