Skip to main content

Table 3 Comparison and agreement of pH and specific gravity measures between the visual vs machine read-out, stratified by infection intensity

From: Evaluation of the Clinitek®, a point-of-care urinalysis system for the measurement of clinically significant urinary metabolites and detection of haematuria in Schistosoma haematobium infected children in southern Côte d’Ivoire

Parameter

Infection intensity

Mean ± SD from visual read

Mean ± SD from machine read

Mean bias (95% CI)

Pearson’s r

pH

Combined

7.31 ± 0.97

7.44 ± 1.03

− 0.14 (− 0.20, − 0.08)

0.88

Light infection

7.37 ± 0.91

7.53 ± 1.00

− 0.16 (− 0.21, − 0.10)

0.89

Heavy infection

6.94 ± 1.16

6.99 ± 1.05

− 0.05 (− 0.24, 0.15)

0.80

Specific gravity

Combined

1.013 ± 0.007

1.018 ± 0.005

− 0.004 (− 0.005, − 0.004)

0.72

Light infection

1.013 ± 0.007

1.017 ± 0.005

− 0.004 (− 0.005, − 0.004)

0.71

Heavy infection

1.015 ± 0.007

1.020 ± 0.006

− 0.005 (− 0.006, − 0.003)

0.70