Skip to main content

Table 3 Combinations of parameters to increase diagnostic accuracy in modeling dataset

From: Identification of parameters and formulation of a statistical and machine learning model to identify Babesia canis infections in dogs using available ADVIA hematology analyzer data

 

Sensitivity

Specificity

LR+

PLT < 102 and PMDW > 1.09

96.6

93.0

13.80

PLT < 102 and MPV > 14

94.3

94.3

16.54

PLT < 102 and %Luc > 1.8

89.7

97.7

39.00

PLT < 102 and PMDW > 1.09 and MPV > 14

93.1

94.8

17.90

PLT < 102 and PMDW > 1.09 and %Luc > 1.8

88.5

98.1

46.58

PLT < 102 and MPV > 14 and %Luc > 1.8

87.4

98.6

62.43