Skip to main content

Table 2 Results of parasite detection of 800 preserved stool samples using FA-280 feces analyzer (user audit) and the formalin ethyl acetate concentration technique (FECT)

From: High-throughput detection of parasites and ova in stool using the fully automatic digital feces analyzer, orienter model fa280

Methods

Number of stool sample

Positive

Negative

Formalin ethyl acetate concentration (FECT)

468

332

FA280 feces analyzer (user audit)

440

360

  1. *Agreement for all parasite positive samples detected between the autoanalyzer using result obtained from user audit and FECT revealed weighted kappa (κ) values of 0.857, indicating almost perfect agreement